Select the baud rate that the computer should use to communicate over the
COM port with the thermometer readout using the Baud rate drop-down list.
The baud rate setting of the thermometer readout must be set to the same value
as this setting. For best results, use the maximum allowable baud rate setting
when possible.
After the model, COM port and baud rate settings are configured, click the Test
Communication button to make sure LogWare is able to communicate with the
thermometer readout. At this time, the thermometer readout must be connected
to the COM port on the computer using a null modem cable. For more informa
-
tion, please refer to the thermometer readout’s User's Guide.
When testing communications with the thermometer readout, there are three
possible results:
The thermometer readout is communicating at the correct baud rate on the
selected COM port.
The thermometer readout is communicating, but possibly at the wrong
baud rate or the thermometer readout model does not match the selected
model.
The thermometer readout failed to communicate at all. Check the COM
port and baud rate settings, make sure the cable is a null modem cable,
and make sure the thermometer readout is plugged in and turned on.
Optional: The Thermometer Settings button may be clicked at this time to gain
access to various thermometer readout settings as applicable to the selected
thermometer readout. This feature is provided solely as a convenience to check
the thermometer readout configuration and/or to configure each channel of the
thermometer readout with appropriate conversion settings and probe coeffi-
cients. The Test Communication button should be clicked prior to clicking the
Thermometer Settings button to make sure the thermometer readout is commu
-
nicating successfully.
Warning: When using the Thermometer Settings tool, make sure that the
thermometer readout is not currently in any menu. Using the thermometer
readout's menus while using the Thermometer Settings tool may cause the
thermometer readout to fail to accept new settings.
Note: When starting a log session, LogWare will override many of the
current thermometer readout settings as needed to acquire data from the
configured channels using the specified log session settings!!
Note: If the serial number and session temperature scale settings have not
been configured prior to clicking the Thermometer Settings button, these
settings will be configured automatically after successfully viewing the
thermometer readout settings dialog.