4.4.2.7.1 Probe tab
Select the Probe tab to view and/or modify probe settings.
Enter the serial number of the probe attached to this channel in the Probe serial
number box. The serial number can be any combination of numbers and/or let-
ters up to 8 characters in length. The underscore “_” character can also be used
in serial numbers.
Select the type of reference junction compensation to use on this channel using
the RJC type drop-down list. Available choices are Internal and External. If Ex-
ternal is chosen as the RJC type, the external RJC temperature must be entered
in the RJC temperature box.
Enter the external RJC temperature value in the RJC temperature box. This box
is disabled if the RJC type is set to Internal.
Note: The RJC temperature value must always be in °C!
Select the conversion type to use for the probe on this channel. The Model
1529 Chub-E4 Thermometer readout supports the following conversion types:
Polynomial
Thermocouple Type (B, E, J, K, N, R, S, T or AuPt)
Volts
Refer to the Report of Calibration for the probe for information on the conver
-
sion type that the probe was calibrated for.
If the selected conversion type requires coefficients or other settings, the Con
-
version Settings button is enabled. Click the Conversion Settings button to dis
-
play the dialog for the selected conversion type. Otherwise, the Conversion
95
4 Tools
Thermometer settings dialogs
Figure 65 1529 Chub-E4 thermocouple channel
settings - Probe tab