The settings on the Calibration tab may be passcode protected. If they are, a
message appears at the bottom of the Calibration tab and the Edit button is dis
-
played. The other controls on this tab are disabled. To edit the calibration pa
-
rameters, click the Edit button and enter the current passcode for the
thermometer readout. If the correct passcode is entered, the calibration parame
-
ter controls are enabled and the values may be changed. The controls remain
enabled until this dialog is closed or until the current changes are applied by
clicking the Apply button.
Caution: Do not change these values unless this module has been
recalibrated or unless these values do not match the values located on the
latest Report of Calibration for this module! Changing these values voids
the current calibration!
Select the date this module was last calibrated using the drop-down calendar or
by entering the date. This date should be located on the Report of Calibration
for this module.
Enter the values for the calibration constants in the boxes provided. The cali-
bration constants required by the 2564 module are: 0 ADJ, 10K ADJ and 100K
ADJ. The values of these constants should be located on the Report of Calibra-
tion for this module.
4.4.3.10 2565 PTC Module Settings dialog
The 2565 Module Settings dialog displays information about the 2565 module
including model number, serial number, identification information and the