Enter the minimum operating temperature for this probe in the MINOP box. If
the temperature measured by the thermometer readout is less than this setting,
the display indicates an overload condition (OL).
Enter the maximum operating temperature for this probe in the MAXOP box. If
the temperature measured by the thermometer readout is greater than this set
-
ting, the display indicates an overload condition (OL).
To exclusively lock the connected probe to this channel, check the Exclusively
lock probe to channel checkbox. Locking a probe to a channel prevents any
other probe from being used on this channel, and also prevents this probe from
being used on any other channel, including channels of other thermometer
readouts.
4.4.1.6 1523/1524 thermocouple channel settings dialog
The Channel Settings dialog for thermocouple probes displays information
about the selected channel.
Some or all of the settings on this dialog may be passcode protected. If so, a
message appears at the bottom of the Probe tab and the Edit button is dis
-
played. The protected controls are disabled. To edit the protected settings, click
the Edit button and enter the current passcode for the thermometer readout. If
the correct passcode is entered, the protected controls are enabled and the val
-
ues may be changed. The controls remain enabled until this dialog is closed or
until the current changes are applied by clicking the Apply button.
After modifying any of the settings on this dialog, click the Apply or OK but
-
ton to write the new settings to the thermometer readout. If the Apply button is
81