2. The Select Model dialog is displayed. Select the thermometer readout
model. Also, select the COM port to which the thermometer readout is
connected and the baud rate to use. Make sure LogWare can communi
-
cate with the thermometer readout by clicking the Test Communication
button. Click the OK button when communication is successful.
3. LogWare attempts to set the thermometer readout’s clock to match the
clock on the computer.
Please refer to Section 4.1.1, Synchronize Date/Time, for more information.
6.8 Protecting calibration parameters and probe
conversion settings
Some of the thermometer readouts supported by LogWare have built-in
passcode protection for calibration parameters and/or probe conversion settings
(including probe characterization coefficients). These settings are typically pro
-
tected because altering their values changes the calibration and/or readout accu
-
racy of the thermometer readout. Some thermometer readouts, however, lack
this protection.
LogWare includes a feature that allows these settings to be passcode protected
(for thermometer readouts that do not inherently include this protection) using
an administrator passcode to prevent unauthorized users from changing the cur-
rent values.
To protect calibration parameters and probe conversion settings:
1. Select the Options option in the Tools menu.
2. The LogWare Options dialog is displayed.
3. Select the Tools tab. The Tools tab contains options that pertain to the
thermometer settings dialogs.
4. In the Passcode Protection section there are two options:
Protect probe conversion settings
Protect calibration parameters
Note: If these settings are disabled, an administrator passcode has been
setup. The administrator passcode is required to change these settings. For
more information on gaining access to the options protected by the admin
-
istrator passcode, see Section 6.9, Setting up an administrator passcode.