4.4.1.3 Calibration tab
Select the Calibration tab to view and/or modify the thermometer readout’s cal
-
ibration settings.
CAUTION!! DO NOT CHANGE THESE VALUES UNLESS THIS READ-
OUT HAS BEEN RECALIBRATED OR UNLESS THESE VALUES DO
NOT MATCH THE VALUES LOCATED ON THE LATEST REPORT OF
CALIBRATION! CHANGING THESE VALUES VOIDS THE CURRENT
CALIBRATION!
Select the date this readout was last calibrated using the drop-down calendar or
by entering the date. This date should be located on the Report of Calibration.
Enter the reference and offset values for each calibration range in the grid. The
calibration ranges are indicated in the first row of the grid. Each calibration
range can have up to 4 adjustment pairs consisting of a reference and an offset
value. Reference and offset values for unused adjustment pairs must be left
blank. These values can be found on the Report of Calibration for the readout
only if this readout received an accredited calibration.