To configure a thermometer readout to read different units on each
channel:
1. Select the Setup option in the Datalog menu to display the Log Session
Settings dialog.
2. Select the Thermometer tab.
3. Select the thermometer readout model using the Model drop-down list.
4. Select the COM port, and baud rate using the drop-down lists in the
Communication settings section.
5. Click the Thermometer Settings button to display the Thermometer Set
-
tings dialog for the selected thermometer readout.
6. On the Thermometer Settings dialog, select the desired system units us
-
ing the Units drop-down list. Click the Apply button to write the new
system units setting to the thermometer readout.
Note: If a non-temperature setting is selected for the system units (i.e.
Ohms, Volts, Ratio, or W), this setting will override any conversion type
setting for all channels that would otherwise cause reading to be dis-
played in temperature!! Therefore, it is recommended that the system units
be set to a temperature setting (i.e. °C, °F, K, etc.) and the conversion type
be used to configure channels that are to read non-temperature values.
7. On the Thermometer Settings dialog, navigate to the first channel to
configure.
a. For thermometer readouts that support modules, select the Modules tab,
select the module from the Modules list that the desired channel is on and
click the Module Settings button. The Module Settings dialog is displayed.
Continue with step b. below.
b. Select the Channels tab, select the desired channel from the Channels
list and click the Channel Settings button. The Channel Settings dialog is
displayed.
8. To configure the selected channel to read temperature, set the Conversion
type to the desired setting and enter the coefficients to use (as applicable)
to convert the readings on that channel to temperature by clicking the
Conversion Settings button.
9. To configure the selected channel to read resistance, voltage, ratio or W,
set the Conversion type appropriately. Refer to the thermometer read
-
out’s User’s Guide for more details on the conversion types supported by
the thermometer readout.
10. Click the OK button on the Channel Settings dialog to write the new set
-
tings to the thermometer readout.
11. Repeat steps 3 to 6 for the other channels that need to be configured.