4.4.2.6.2 Calibration tab
Select the Calibration tab to view and/or modify channel calibration informa
-
tion.
Caution: Do not change these values unless this channel has been
recalibrated or unless these values do not match the values located on the
latest Report of Calibration for this channel! Changing these values voids
the current calibration!
Select the date this channel was last calibrated using the drop-down calendar or
by entering the date. This date should be located on the Report of Calibration
for this channel.
Select the date this channel is due for recalibration using the drop-down calen
-
dar or by entering the date.
Enter the values for the calibration constants in the boxes provided. The cali
-
bration constants required by a PRT/Thermistor channel are: C0, C100, C400,
C0K, C10K and C100K. The values of these constants should be located on the
Report of Calibration for this channel.
93
4 Tools
Thermometer settings dialogs
Figure 63 1529 Chub-E4 PRT/thermistor channel
settings - Calibration tab