If the sample action is set to Run or Stop, the Sample n setting is not applicable
and is disabled. If the filter type is set to Sample n,theSamplen setting is
enabled.
If the sample action is set to Sample n, enter the number of samples in the Sam
-
ple n box. The Sample n setting can be set to any value between 1 and 32,000.
The Sample timing controls allow the thermometer readout’s sample timing set
-
tings to be viewed and/or modified. The sample timing settings include the
Sample interval, Integration time and Conversion time.
Enter the sample interval value in the Sample interval box provided or use the
up/down buttons. The sample interval can be set to any value between 2 and
99,999 seconds.
Note: The sample interval must be greater than or equal to the integration
time, which in turn must be greater than or equal to the conversion time. If
the sample interval is set to a value that is less than the current integration
time, the integration time and conversion time settings are automatically
set to match the sample interval setting!
Enter the integration time value in the Integration time box provided or use the
up/down buttons. The integration time can be set to any value between 2 and
9,999 seconds.
Note: The integration time must be greater than or equal to the conver
-
sion time. If the integration time is set to a value that is less than the cur
-
rent conversion time, the conversion time setting is automatically set to
match the integration time setting!
Enter the conversion time value in the Conversion time box provided or use the
up/down buttons. The conversion time can be set to any value between 2 and 10
seconds.
The Resolutions controls allow the thermometer readout’s resolution settings to
be viewed and/or modified. The resolution settings include the Temperature,
Resistance and Ratio settings.
Enter the resolution for temperature readings in the Temperature box provided
or use the up/down buttons. The temperature resolution can be set to any value
between 0 and 8 digits (places after decimal point).