4.4.3.15.1 Probe tab
Select the Probe tab to view and/or modify probe settings.
The settings on the Probe tab may be passcode protected. If they are, a message
appears at the bottom of the Probe tab and the Edit button is displayed. The
other controls on this tab are disabled except the Conversion Settings button. If
the selected conversion type does not have any settings, this button should also
be disabled. To edit the probe settings, click the Edit button and enter the cur-
rent passcode for the thermometer readout. If the correct passcode is entered,
the probe setting controls are enabled and the values may be changed. The con-
trols remain enabled until this dialog is closed or until the current changes are
applied by clicking the Apply button.
Enter the serial number of the probe attached to this channel in the probe serial
number box. The serial number can be any combination of numbers and/or let
-
ters up to 8 characters in length. The dash “-“ character can also be used in se
-
rial numbers.
Select the type of cold junction compensation to use on the selected channel us
-
ing the CJC type drop-down list. Available choices are Internal and External. If
External is chosen as the CJC type, the external CJC temperature must be en
-
tered in the CJC temperature box.
Enter the external CJC temperature value in the CJC temperature box. This box
is disabled if the CJC type is set to Internal.
Note: The CJC temperature value must always be in °C!
Select the conversion type to use for the probe on this channel. Available con
-
version types will vary depending on the module that this channel belongs to.
Refer to the Report of Calibration for the probe attached to this channel for in
-
formation on the conversion type that the probe was calibrated for.
User’s Guide
136
Figure 111 1560 Black Stack PRT/Thermistor
Channel Settings - Probe tab