clicked, this dialog remains open. If the OK button is clicked, this dialog closes
after writing the settings.
To close this dialog without writing the channel settings to the thermometer
readout, click the Cancel button.
Click the Help button to open the appropriate help topic for this dialog.
Use the Type drop-down list to select the probe type for the selected probe.
Possible probe types are PRT, Thermistor or Resistor.
The Calibration Date may be used to record the date the probe was calibrated.
The calibration date is an optional parameter. The checkbox next to the calibra
-
tion date field indicates whether the probe's calibration date is saved in the in
-
strument. If the checkbox is checked, the calibration date is saved in the
instrument. If the checkbox is unchecked, the probe's calibration date is set to
zeros in the instrument.
The Nominal Current is the usual sensing current that should be used to mea
-
sure the resistance of the probe. Possible values depend on the probe type
selected.
PRT probe can be set to any value between 0.01 and 20.0 mA
Thermistor probe can be set to any value between 0.001 and 0.1 mA
Resistor probe can be set to any value between 0.001 and 20 mA.
The Maximum Current sets a limit for the probe's sensing current. Possible val-
ues depend on the probe type selected.
PRT probe can be set to any value between 0.01 and 20.0 mA
Thermistor probe can be set to any value between 0.001 and 0.1 mA
Resistor probe can be set to any value between 0.001 and 20 mA.
The Maximum Temperature sets the temperature limit for the probe. Possible
values depend on the probe type selected.
PRT probe can be set to any value between 25.0 and 2000.0 °C
Thermistor probe can be set to any value between 25.0 and 1000.0 °C
Maximum temperature is not available for a resistor probe.
Select the conversion type to use for the selected probe. The Model
1594A/1595A Hart Super-Thermometer readout supports the following conver
-
sion types:
ITS-90
CVD-ALPHA
CVD-ABC
PT-100
R(T)
T(R)
167