2.3.2 Log Session tab
After the thermometer readout and communication settings are configured, the
log session settings may be configured.
Select the interval at which LogWa re should take readings on subsequent chan-
nels using the Interval between readings settings. The range of possible settings
is limited by the capabilities of the selected thermometer readout. For most
Hart Scientific multi-channel digital thermometer readouts, the minimum sam-
ple interval is 2 seconds.
Note: Due to the latency inherent in the measurement process, a thermom
-
eter readout may not be able to take readings at the exact selected interval
(i.e. the 1560 thermometer readout allows a minimum interval of 2 sec
-
onds, but actually takes about 2.3 seconds to make one measurement). For
this reason, using the minimum interval supported by the thermometer
readout may cause some duplicate readings to be taken. To avoid this
problem, see the Wait until new measurement is ready (when applicable)
option on the LogWare Options dialog. Please refer to Seciton 5.6.4, Op
-
tions, for more details.
If this interval is set to a value of 0, LogWare will attempt to take simultaneous
readings on all enabled channels of the thermometer readout. If this feature is
used, you should use the Interval between scan sequences setting to limit how
often all channels are sampled.
Note: when using this feature, the Wait until new measurement is ready
option on the General tab of the Options dialog MUST be disabled (un
-
checked).
User’s Guide
22
Figure 11 Log Session Settings - Log Session tab