4.4.3.3 Measurement tab
Select the Measurement tab to view and/or modify measurement related set
-
tings.
Select the scan mode for the 1560 Black Stack Thermometer readout using the
Scan mode drop-down list. The scan mode can be set to one of the following
options:
Primary channel only
Alternate scan/primary channels
Scan channels only
If the Primary channel only option is selected, the thermometer readout will be
set to measure only the channel selected in the Primary channel drop-down list
below.
If the Alternate scan/primary channels option is selected, the thermometer read
-
out will be set to alternate measurements on the primary channel and each of
the enabled scan channels.
If the Scan channels only option is selected, the thermometer readout will be
set to scan through each of the enabled channels in the Scan channels list.
Select the channel to set as the primary channel using the Primary channel
drop-down list. This list indicates the channel number, module name and mod
-
ule input number for all channels on the thermometer readout. The primary
channel setting is only enabled if the scan mode setting is set to the Primary
channel only or the Alternate scan/primary channels option.
Select the channels to include in the thermometer readout’s scan sequence. The
Scan channels list indicates the channel number, module name and module in
-
put number for all channels on the thermometer readout. Each channel can be
enabled or disabled by checking or unchecking the box next to the channel
number independent of all other channels. The Scan channels list is only en
-
User’s Guide
100
Figure 70 1560 Black Stack settings - Measurement tab