4.4.6.3 4.4.6.3 Timing tab
Select the Timing tab to view and/or modify timing settings.
Use the Sample Period drop-down list to select the length of time required to
obtain an individual measurement. Possible values are 1 s, 2 s, 5 s and 10 s.
Note: Changing this setting may affect the Display Interval setting, as
Sample Period cannot be greater than Display Interval.
Use the Display Interval drop-down list to select how often the instrument's dis
-
play, statistics fields, memory and communications ports are updated. Possible
valuesare1s,2s,10s,30sand60s.
Note: Changing this setting may affect the Sample Period setting, as Sam
-
ple Period cannot be greater than Display Interval.
Use the Average Enable drop-down list to select whether averaging is enabled
(On) or disabled (Off). If the setting is On, all measurements acquired at the
Sample Period rate are averaged prior to updating the instrument's display. If
the setting is Off, the latest measurement acquired at the Sample Period rate is
used to update the instrument's display.
Use the Stop Timer Enable drop-down list to select whether the stop timer is
enabled (On) or disabled (Off). If the setting is On, a timer begins counting
down when a measurement is started and stops the measurement automatically
163
4 Tools
Thermometer settings dialogs
Figure 131 1594A/1595A Hart Super-Thermometer Timing tab