Use the Filter type drop-down list to select the type of filtering to use. The filter
type can be set to one of the following options:
None
Average
Exponential
If the filter type is set to None, the Response time and Threshold settings are
not applicable and are disabled. If the filter type is set to Average or Exponen
-
tial, the Response time and Threshold settings are enabled.
Enter the response time for the filter in the Response time box. The response
time can be set to any value between 2 and 200 seconds.
Enter the threshold to use for the selected filter in the Threshold box. The
threshold can be set to any value between .0001 and 1,000,000.
4.4.5.2 Sampling tab
Select the Sampling tab to view and/or modify sampling settings.
Select the units using the Units drop-down list. Readings can be displayed in
either °C, °F, K, Ohms, Ratio or W.
Note: This setting may override the selected conversion type for each
channel. Please refer to the thermometer readout’s User’s Guide for more
information on setting the system units.
Select the sample action from the Sample action drop-down list. The sample
action can be set to one of the following options:
151
4 Tools
Thermometer settings dialogs
Figure 122 1590 Super-Thermometer II Settings -
Sampling tab