in from the thermometer readout. The Channel Settings dialog is described in
detail in Section 4.4.3.15, 1560 Black Stack Thermocouple Channel Settings
Dialog.
4.4.3.11.2 Calibration tab
Select the Calibration tab to view and/or modify module calibration informa
-
tion.
The settings on the Calibration tab may be passcode protected. If they are, a
message appears at the bottom of the Calibration tab and the Edit button is dis
-
played. The other controls on this tab are disabled. To edit the calibration pa
-
rameters, click the Edit button and enter the current passcode for the
thermometer readout. If the correct passcode is entered, the calibration parame
-
ter controls are enabled and the values may be changed. The controls remain
enabled until this dialog is closed or until the current changes are applied by
clicking the Apply button.
Caution: Do not change these values unless this module has been
recalibrated or unless these values do not match the values located on the
latest Report of Calibration for this module! Changing these values voids
the current calibration!
The AMP GA constant should always be set to 1 and cannot be changed.
User’s Guide
124