when the timer elapses. If the setting is Off, the Timer Duration is not applica
-
ble and is disabled.
The Timer Duration determines the amount of time a measurement runs. If the
Stop Timer Enable is Off, this setting is not applicable and is disabled. The
timer duration can be set to any time between 00:00:01 and 23:59:59 with the
time format set as hh:mm:ss.
4.4.6.4 Scanning tab
Select the Scanning tab to view and/or modify scan settings.
Use the Scan Enable drop-down list to select whether scanning is enabled (On)
or disabled (Off). If Scan Enable is Off, Repeat Count, Alternate Reference and
Reference Channel are not applicable and are disabled and only one channel
can be selected in the Scan Channels list.
The Repeat Count determines how many times a channel is measured before
the instrument continues to the next channel to be measured. If Scan Enable is
Off, this setting is not applicable and is disabled.
Use the Alternate Reference drop-down list to select whether the reference
channel is automatically measured after each individual channel is measured. If
this setting is Off, the Reference Channel is not applicable and is disabled. If
Scan Enable is Off, this setting is not applicable and is disabled.
When the Alternate Reference is enabled (On), this setting determines which
channel shall be used for the reference. If Scan Enable or Alternate Reference
is Off, this setting is not applicable and is disabled.
User’s Guide
164
Figure 132 1594A/1595A Hart Super-Thermometer Scanning tab