The Resistor Value is the calibrated resistance of the resistor.
The Maximum Current sets a limit for the resistor's sensing current. Possible
values are any value between 0.01 and 20.0 mA.
The Calibration Date may be used to record the date the resistor was calibrated.
The calibration date is an optional parameter. The checkbox next to the calibra
-
tion date field indicates whether the resistor's calibration date is saved in the in
-
strument. If the checkbox is checked, the resistor's calibration date is saved in
the instrument. If the checkbox is unchecked, the resistor's calibration date is
set to zeros in the instrument.
The Calibration Due date may be used to record the date the resistor is due for
recalibration. The calibration due date is an optional parameter. The checkbox
next to this field indicates whether the resistor's calibration due date is saved in
the instrument. If the checkbox is checked, the resistor's calibration due date is
saved in the instrument. If the checkbox is unchecked, the resistor's calibration
due date is set to zeros in the instrument.
171
4 Tools
Thermometer settings dialogs