9.7 W(T90) Settings dialog
The W(T90) Settings dialog allows the RTPW coefficient to be entered.
If the W(T90) coefficients were read in from the thermometer readout, the
boxes are filled with the current coefficient values.
Enter the RTPW coefficient from the Report of Calibration in the RTPW box.
This box must contain a valid value.
Select the Open button to display the Windows® Open dialog to open a previ-
ously saved configuration file which contains W(T90) coefficients. Select the
Save As button to display the Windows® Save As dialog to save the current
W(T90) coefficients to a configuration file.
Click the OK button to accept the current coefficients and close this dialog.
Click the Cancel button to discard any changes you have made and close this
dialog.
Click the Help button to display the appropriate help topic for this dialog.
9.8 Thermistor Settings dialog
The Thermistor Settings dialog allows the Thermistor T(R) or Thermistor R(T)
coefficients to be entered.
If the Thermistor coefficients were read in from the thermometer readout, the
boxes are filled with the current coefficient values.
Select the Open button to display the Windows® Open dialog to open a previ
-
ously saved configuration file which contains Thermistor coefficients. Select
the Save As button to display the Windows® Save As dialog to save the current
Thermistor coefficients to a configuration file.
247
9 Conversion settings dialogs
W(T90) Settings dialog
Figure 178 W(T90) Settings dialog