Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 90

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Instrument setup
R&S
®
MXO 4 Series
90User Manual 1335.5337.02 ─ 05
Start alignment
Starts the self-alignment procedure for all channels.
Date, Time, Overall alignment state
Show the date, time and the summary result of the self-alignment process: not aligned,
passed or failed. Detailed results are provided on the "Alignment results" tab.
Remote command:
CALibration:DATE? on page 424
CALibration:TIME? on page 424
CALibration:RESult? on page 425
Show results
Opens a dialog to display the alignment results.
For each channel, the results of the individual alignment steps are shown for all techni-
cal channel component. In case you require support, you may be asked to provide this
information.
Maintenance settings
Loading ...
Loading ...