Instrument setup
R&S
®
MXO 4 Series
88User Manual 1335.5337.02 ─ 05
For more information, refer also to "R&S MXO 4 instrument security procedures". It is
available for download on the Rohde & Schwarz webpage.
5.6.4 Restoring settings
When you have changed many settings on the instrument, and you are not sure which
settings are causing which effect, you can restore the default settings and start anew.
The following methods are available:
Saving instrument settings as a user-defined preset and restoring the instrument
settings to user-defined default values
Restoring all settings on the R&S MXO 4 to the factory-defined values
Restoring settings from a file
To save a user-defined preset
1. Open the "Menu" > "Settings" > "Preset" tab.
2. Enter a name for the preset file. Select the file format.
3. Tap "Save preset file".
Note: If you want to store the file in another directory than the displayed one,
select "Save as". See also: Chapter 12.5, "File browser dialog", on page 273.
To restore the instrument settings to user-defined default values
1. Open the "Menu" > "Settings" > "Preset" > "Settings" tab.
2. In "Preset file name", enter the name of the file that contains the required settings.
3. To use these settings as preset values, select "Enable user-defined preset".
4. Press the [Preset] key.
To restore all settings to the factory defaults
1. Open the "Menu" > "Settings" > "Preset" > "Factory" tab.
2. Tap the "Factory preset" button.
All settings on the R&S MXO 4 are reset to their factory-defined values.
Preset setup