Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 425

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
425User Manual 1335.5337.02 ─ 05
Return values:
<Time>
Usage: Query only
Asynchronous command
Manual operation: See "Date, Time, Overall alignment state" on page 90
CALibration:RESult?
Returns the result of the last selfalignment and the current alignment status. In remote
mode, *CAL? provides more detailed information.
Return values:
<ResultState> PASSed | FAILed | NOALigndata
*RST: FAILed
Usage: Query only
Asynchronous command
Manual operation: See "Date, Time, Overall alignment state" on page 90
SYSTem:APUP <AutoPowerUp>
If enabled, the instrument powers up automatically when it is connected to the mains
voltage.
Parameters:
<AutoPowerUp> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Auto power up" on page 92
17.8 Acquisition and setup
17.8.1 Starting and stopping acquisition
RUNCont...................................................................................................................... 425
RUN............................................................................................................................. 425
RUNSingle.................................................................................................................... 426
SINGle..........................................................................................................................426
STOP........................................................................................................................... 426
RUNCont
RUN
Starts the continuous acquisition.
Acquisition and setup
Loading ...
Loading ...
Loading ...