Remote control commands
R&S
®
MXO 4 Series
589User Manual 1335.5337.02 ─ 05
17.16.1 Configuration settings for all serial protocols
17.16.1.1 General settings
SBUS<sb>[:STATe]........................................................................................................ 589
SBUS<sb>:TYPE...........................................................................................................589
SBUS<sb>:RESult......................................................................................................... 589
SBUS<sb>:FORMat....................................................................................................... 590
SBUS<m>:ZCOupling.....................................................................................................590
SBUS<m>:THReshold....................................................................................................590
SBUS<m>:SETReflevels................................................................................................ 590
SBUS<sb>[:STATe] <ProtocolState>
Enables the decoding of the specified bus.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<ProtocolState> ON | OFF
Usage: Asynchronous command
SBUS<sb>:TYPE <Protocol Type>
Selects the bus type for analysis. The type of available buses depends on the installed
options.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Protocol Type> I2C | SPI | UART | CAN
Usage: Asynchronous command
Manual operation: See "Protocol type" on page 276
SBUS<sb>:RESult <ShwResTbl>
Enables a table with decoded data of the serial signal. The function requires the option
for the analyzed protocol.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<ShwResTbl> ON | OFF
*RST: OFF
Usage: Asynchronous command