Waveform generator (option R&S MXO4-B6)
R&S
®
MXO 4 Series
371User Manual 1335.5337.02 ─ 05
Arb wfm source
Selects the source of the arbitrary waveform.
You can load CSV files that follow a defined structure, or saved reference waveforms.
See also: "Open" on page 371.
Remote command:
WGENerator<wg>:ARBGen[:SOURce] on page 716
Sample rate
Sets the sample rate for the arbitrary waveform.
If the sample rate is given in the arbitrary waveform file (CSV or REF file), the instru-
ment displays the value here, and you can change it. If no sample rate is available in
the data, you can set the required sample rate.
Remote command:
WGENerator<wg>:ARBGen:SRATe on page 716
Number of samples
Displays the number of samples in the loaded waveform.
Remote command:
WGENerator<wg>:ARBGen:SAMPles? on page 715
Open
Opens a dialog to select the saved waveform, and loads the file. The setting is availa-
ble if "Arb wfm source" is set to "Saved waveform".
You can load CSV files, or REF waveforms (saved reference waveforms). See also:
Chapter 15.1.4.1, "Arbitrary waveform files", on page 368.
When the file is loaded, the output voltage is set to the value of the first sample in the
file. Therefore, if you want to output single periods of the arbitrary waveform, first load
the waveform file, and then connect the DUT.