Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 472

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
472User Manual 1335.5337.02 ─ 05
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
HIGH = stays high, the signal level stays above the trigger level.
LOW = stays low, the signal level stays below the trigger level.
EITHer = stays high or low.
*RST: HIGH
Usage: Asynchronous command
Manual operation: See "Range" on page 149
TRIGger:EVENt<ev>:TIMeout:TIME
<Time>
Sets the time limit for the timeout at which the instrument triggers.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Time> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 1E-07
Default unit: s
Usage: Asynchronous command
Manual operation: See "Time" on page 149
17.9.9 Interval trigger
TRIGger:EVENt<ev>:INTerval:DELTa...............................................................................472
TRIGger:EVENt<ev>:INTerval:RANGe............................................................................. 473
TRIGger:EVENt<ev>:INTerval:SLOPe..............................................................................473
TRIGger:EVENt<ev>:INTerval:WIDTh.............................................................................. 473
TRIGger:EVENt<ev>:INTerval:DELTa <WidthDelta>
Sets a range around the interval width value specified with TRIGger:EVENt<ev>:
INTerval:WIDTh.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<WidthDelta> Range: 0 to 10
Increment: 1E-07
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "±Delta" on page 151
Trigger
Loading ...
Loading ...
Loading ...