Remote control commands
R&S
®
MXO 4 Series
476User Manual 1335.5337.02 ─ 05
Usage: Asynchronous command
Manual operation: See "Range" on page 153
TRIGger:EVENt<ev>:SLEW:RATE <Time>
For TRIGger:EVENt<ev>:SLEW:RANGe = INSRange and OUTRange, the slew rate
defines the center of a range which is defined by the limits "±Delta".
For TRIGger:EVENt<ev>:SLEW:RANGe = LTHan and GTHan, the slew rate defines
the maximum and minimum slew rate limits, respectively. When the signal crosses this
level, the slew rate measurement starts or stops depending on the selected slope (see
TRIGger:EVENt<ev>:SLEW:SLOPe).
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Time> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 1E-10
Default unit: s
Usage: Asynchronous command
Manual operation: See "Slew rate" on page 153
TRIGger:EVENt<ev>:SLEW:SLOPe <Slope>
Sets the edge direction for the trigger.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Slope" on page 152
17.9.11 Setup & Hold
TRIGger:EVENt<ev>:SETHold:CSOurce:EDGE................................................................476
TRIGger:EVENt<ev>:SETHold:CSOurce:LEVel.................................................................477
TRIGger:EVENt<ev>:SETHold:CSOurce[:VALue]..............................................................477
TRIGger:EVENt<ev>:SETHold:HTIMe..............................................................................477
TRIGger:EVENt<ev>:SETHold:STIMe..............................................................................478
TRIGger:EVENt<ev>:SETHold:CSOurce:EDGE <ClockEdge>
Sets the edge of the clock signal. Edge and level define the time reference point.