Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 709

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
709User Manual 1335.5337.02 ─ 05
Parameters:
<PeriodCarrierAlias> Range: 1E-08 to 1000
Increment: 1
*RST: 1E-06
Default unit: s
Usage: Asynchronous command
WGENerator<wg>:MODulation:FM:DCYCle <SquareDutyCycle>
Sets the duty cycle for a square waveform. The duty cycle expresses for what percent-
age fraction of the period, the waveform is active, i.e. the signal state is high.
Suffix:
<wg>
.
1..2
Parameters:
<SquareDutyCycle> Range: 10 to 90
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Duty cycle" on page 366
WGENerator<wg>:MODulation:FM:DEViation <Deviation>
Sets the frequency deviation, the maximum difference between the FM modulated sig-
nal and the carrier signal.
Suffix:
<wg>
.
1..2
Parameters:
<Deviation> Range: 0.001 to 10000000
Increment: 1
*RST: 1000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Deviation" on page 366
WGENerator<wg>:MODulation:FM:FREQuency <Frequency>
Sets the frequency of the modulating waveform for FM modulation.
Suffix:
<wg>
.
1..2
Waveform generator (option R&S MXO4-B6)
Loading ...
Loading ...
Loading ...