Remote control commands
R&S
®
MXO 4 Series
643User Manual 1335.5337.02 ─ 05
Parameters:
<Bitrate> Range: 10000 to 5000000
Increment: 100
*RST: 100000
Default unit: bps
Usage: Asynchronous command
Manual operation: See "Nominal bit rate" on page 319
SBUS<sb>:CAN:SAMPlepoint <SamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<SamplePoint> Range: 12 to 96
Increment: 5
*RST: 66
Default unit: %
Usage: Asynchronous command
Manual operation: See "Sample point" on page 319
SBUS<sb>:CAN:DATA:HYSTeresis <Hysteresis>
Sets a hysteresis value for the data line.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Hysteresis>
Usage: Asynchronous command
SBUS<sb>:CAN:DATA:SOURce DATASource
Sets the source channel to which the data line is connected.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
DATASource C1 | C2 | C3 | C4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9
| D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 |
R1 | R2 | R3 | R4
Usage: Asynchronous command
Manual operation: See "Data" on page 318
Protocols