Remote control commands
R&S
®
MXO 4 Series
489User Manual 1335.5337.02 ─ 05
Parameters:
<Delay> Range: 6E-07 to 1
Increment: 1E-09
*RST: 6E-07
Default unit: s
Usage: Asynchronous command
Manual operation: See "Delay" on page 167
TRIGger:ACTions:OUT:PLENgth <PulseLength>
Sets the length of the trigger out pulse.
Parameters:
<PulseLength> Range: 1.6E-08 to 0.05
Increment: 1.6E-08
*RST: 9.6E-08
Default unit: s
Usage: Asynchronous command
Manual operation: See "Pulse length" on page 167
TRIGger:ACTions:OUT:POLarity <Polarity>
Sets the polarity of the trigger out pulse, which is the direction of the first pulse edge.
Parameters:
<Polarity> POSitive | NEGative
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Polarity" on page 167
17.10 Waveform analysis
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
Waveform analysis