Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 119

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Acquisition and waveform setup
R&S
®
MXO 4 Series
119User Manual 1335.5337.02 ─ 05
PROBe<ch>:SETup:STATe? on page 445
PROBe<ch>:SETup:ATTenuation:DEFProbe on page 442
Probe bandwidth
Shows the bandwidth of the connected probe. For probes that are not detected or pre-
defined, set the bandwidth manually.
Remote command:
PROBe<ch>:SETup:BANDwidth? on page 443
Offset
Channel offset that is also set on "Vertical" > "Setup" tab. See "Offset" on page 113.
Vertical unit
Shows the unit of the connected probe if the probe is detected or predefined. For
unknown probes, select the unit that the probe can measure.
Remote command:
PROBe<ch>:SETup:ATTenuation:UNIT on page 443
Manual attenuation
Shows the attenuation of the connected probe if the probe is detected or predefined.
For unknown probes, set the correct attenuation of the probe.
The vertical scaling and measured values are multiplied by this factor so that the dis-
played values are equal to the actual measured signal values.
Make sure to set the attenuation factor on the instrument according to the probe being
used. Otherwise, the measurement results do not reflect the actual voltage level, and
you might misjudge the actual risk.
Remote command:
PROBe<ch>:SETup:ATTenuation[:AUTO]? on page 442
PROBe<ch>:SETup:ATTenuation:MANual on page 442
Set offset to mean
Compensates automatically for a DC component of the input signal using the result of
a background mean measurement.
The result is shown in "Offset". The function is probe-independent and supports quick
and convenient measurements of input signals with different DC offsets. It detects off-
set values even when the signal is out of the current measurement range. It also sets
the zero level to the determined DC offset in the middle of the screen and thus pre-
vents clipping of the waveform.
Remote command:
PROBe<ch>:SETup:OFFSet:TOMean on page 444
Probes
Loading ...
Loading ...
Loading ...