Remote control commands
R&S
®
MXO 4 Series
448User Manual 1335.5337.02 ─ 05
DIFFerential
Measures differential and common mode voltages
SINGleended
Measures the voltage between the positive/negative signal
socket and the ground.
*RST: DIFFerential
Usage: Asynchronous command
Manual operation: See "Display" on page 124
PROBe<ch>:SETup:ZAXV
<AttenuationZA15>
If you use the external attenuator R&S RT-ZA15 together with one of the differential
active probes R&S RT-ZD10/20/30, enable "RT-ZA15 attenuator" to include the exter-
nal attenuation in the measurements.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<AttenuationZA15> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "RT-ZA15 attenuator" on page 124
PROBe<ch>:PMETer:STATe <State>
Activates the integrated R&S ProbeMeter on probes with Rohde & Schwarz probe
interface.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "ProbeMeter" on page 122
PROBe<ch>:PMETer:RESults:SINGle?
Returns the ProbeMeter measurement result of single-ended active Rohde & Schwarz
probes, the voltage measured between the probe tip and the ground.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Acquisition and setup
Loading ...
Loading ...
Loading ...