Remote control commands
R&S
®
MXO 4 Series
482User Manual 1335.5337.02 ─ 05
Manual operation: See "State sources" on page 157
17.9.14 Trigger mode, holdoff
TRIGger:MODE............................................................................................................. 482
TRIGger:FORCe............................................................................................................482
TRIGger:HOLDoff:MODE................................................................................................483
TRIGger:HOLDoff:AUTotime?..........................................................................................483
TRIGger:HOLDoff:SCALing.............................................................................................484
TRIGger:HOLDoff:EVENts.............................................................................................. 484
TRIGger:HOLDoff:MAX.................................................................................................. 484
TRIGger:HOLDoff:MIN....................................................................................................485
TRIGger:HOLDoff:TIME..................................................................................................485
TRIGger:MODE <TriggerMode>
Sets the trigger mode which determines the behavior of the instrument with and with-
out a trigger event.
Parameters:
<TriggerMode> AUTO | NORMal | FREerun
AUTO
The instrument triggers repeatedly after a time interval if the trig-
ger conditions are not fulfilled. If a real trigger occurs, it takes
precedence. The time interval depends on the time base.
NORMal
The instrument acquires a waveform only if a trigger occurs.
FREerun
The instrument triggers after a very short time interval - faster
than in AUTO mode. Real triggers are ignored.
*RST: AUTO
Usage: Asynchronous command
Manual operation: See "[Auto Norm]" on page 36
TRIGger:FORCe
Provokes an immediate single acquisition. Force the trigger if the acquisition is running
in normal mode and no valid trigger occurs. Thus, you can confirm that a signal is
available and use the waveform display to determine how to trigger on it.
Usage:
Setting only
Asynchronous command
Manual operation: See "Force trigger" on page 162