Acquisition and waveform setup
R&S
®
MXO 4 Series
128User Manual 1335.5337.02 ─ 05
Current probes R&S RT-ZCxxB
Current probes R&S RT-ZCxxB have a Rohde & Schwarz probe interface; they are
powered and remotely controlled by the oscilloscope.
When the probe is connected, demagnetization is performed automatically.
Figure 6-7: Setup for current probe R&S
RT-ZC05B
Most settings are common settings, which are available for all probe types. See Chap-
ter 6.4.1, "Common probe settings", on page 117 for description of these settings.
Current probes are adjusted by the following functions.
Gain, Manual gain
Shows the gain of the connected probe if the probe is detected or predefined. For
unknown current probes, set the correct gain of the probe.