Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 483

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
483User Manual 1335.5337.02 ─ 05
TRIGger:HOLDoff:MODE <Mode>
Selects the method to define the holdoff condition.
The trigger holdoff defines when the next trigger after the current will be recognized.
Thus, it affects the next trigger to occur after the current one. Holdoff helps to obtain
stable triggering when the oscilloscope is triggering on undesired events.
Holdoff settings are not available if the trigger source is an external trigger input or
serial bus, and if you trigger on a sequence of events.
Parameters:
<Mode> TIME | EVENts | RANDom | AUTO | OFF
TIME
Defines the holdoff directly as a time period. The next trigger
occurs only after the holdoff time has passed, which is defined
using TRIGger:HOLDoff:TIME).
EVENts
Defines the holdoff as a number of trigger events. The next trig-
ger occurs only when this number of events is reached. The
number of triggers to be skipped is defined with TRIGger:
HOLDoff:EVENts.
RANDom
Defines the holdoff as a random time limited by TRIGger:
HOLDoff:MIN and TRIGger:HOLDoff:MAX. For each acquisi-
tion cycle, the instrument selects a new random holdoff time
from the specified range.
AUTO
The holdoff time is calculated automatically based on the current
horizontal scale.
OFF
No holdoff
*RST:
OFF
Usage: Asynchronous command
Manual operation: See "Holdoff mode" on page 162
TRIGger:HOLDoff:AUTotime?
Returns the resulting holdoff time, if TRIGger:HOLDoff:MODE is set to AUTO: Auto
time = Auto time scaling * Horizontal scale. The auto time scaling factor is defined with
TRIGger:HOLDoff:SCALing.
Return values:
<AutoTime> Range: 1E-07 to 10
Increment: 0.0002
*RST: 0.001
Default unit: s
Trigger
Loading ...
Loading ...
Loading ...