Waveform generator (option R&S MXO4-B6)
R&S
®
MXO 4 Series
369User Manual 1335.5337.02 ─ 05
Time values are ignored. You can specify just rate and voltage values as below:
Rate = 5000000 //Sample rate of the arbitrary waveform
-5.995 //Voltage value 1
-6.0 //Voltage value 2
.......
Without specified Rate and without specified Time values:
The last user-defined sample rate is used to calculate the waveform. You can
change the "Sample rate" in the user interface. The voltage values are then played
with this sample rate.
-5.995 //Voltage value 1
-6.0 //Voltage value 2
.......
Without specified Rate and with specified Time:
The timing information of the first 2 time values is used to calculate the sample
rate.
Example:
Consider the following file:
0.000000E+000,-5.995 //Time value 1, Voltage value 1
1.237011E-005,-6.0 //Time value 2, Voltage value 2
The sample rate is:
Sample rate = 1/ Time between first two samples =1 / 1.237011E-005 =
80.840KSampe/sec
15.1.4.2 Arbitrary settings
Access: "Gen" signal activator> "Gen1" or "Gen2" > "Setup" tab > "Function type" =
"Arbitrary"
The following settings in the "Setup" tab and the "Arbitrary Setup" dialog are specific
for generated arbitrary waveforms.
Setup of the waveform generator