Network operation and remote control
R&S
®
MXO 4 Series
393User Manual 1335.5337.02 ─ 05
Bit No. Meaning
4
MEASuring
The bit is set as long as an acquisition - sampling and postprocessing - is running. In run contin-
uous mode, the bit is always set.
5 - 15 Not used
16.5.3.4 STATus:QUEStionable register
This register contains specific information on instrument operation and signal reliability.
It can be read using the commands STATus:QUEStionable:CONDition? and
STATus:QUEStionable[:EVENt]?.
The remote commands for the STATus:QUEStionable register are described in Chap-
ter 17.19, "Status reporting", on page 719.
Table 16-5: Bits in the STATus:QUEStionable register
Bit
No.
Meaning
0 - For future use.
1
PPSupply
This bit is set if probe power supply overload occurs.
2
POVerload
This bit is set if a probe overload occurs.
3
COVerload
This bit is set if a questionable channel overload occurs.
4
TEMPerature
This bit is set if a questionable temperature occurs.
5
ADCState
The bit is set if the signal is clipped on the upper or lower edge of the screen-over-
flow of the ADC occurs.
6
GOVerload
This bit is set if a generator overload occurs.
7 - For future use.
8
NOALigndata
This bit is set if no alignment data is available - the instrument is uncalibrated.
9
LIMit
This bit is set if a limit value is violated.
10
MARGin
This bit is set if a margin value is violated, and the limit value is not violated.
11
IMPRecise
This bit is set if the magnitude of the signal is too low to get reliable measurement
results.
12 - For future use.
13 - For future use.
14 - For future use.
16.5.3.5 Application of the status reporting system
The purpose of the status reporting system is to monitor the status of one or several
instruments in a test system. In particular, it monitors continuous measurements to
detect errors in the signal. The controller must receive and evaluate the information of
all devices.
Remote control - status reporting system