Mixed signal option (MSO, R&S MXO4-B1)
R&S
®
MXO 4 Series
354User Manual 1335.5337.02 ─ 05
D0-D15
Sets an individual delay for each digital channel to time-align it with other digital chan-
nels.
The skew value compensates delays that are known from the circuit specifics or
caused by the different length of cables. The skew between the probe boxes of the dig-
ital channels and the probe connectors of the analog channels is automatically aligned
by the instrument.
Remote command:
PBUS<pb>:BIT<n>:SKEW on page 691
Skew
Sets a general delay for all digital channels.
Remote command:
PBUS<pb>:SKEW on page 696