Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 429

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
429User Manual 1335.5337.02 ─ 05
ACQuire:ROLLmode:OSCapture <OffScreenCapture>
If enabled, the analyzable waveform in roll mode is extended. Samples that arrive the
left edge of the display are kept, they are moved to the unvisible area on the left.
Parameters:
<OffScreenCapture> ON | OFF
*RST: OFF
Example: See Chapter 17.4.3, "Data transfer in roll mode", on page 403.
Usage: Asynchronous command
Manual operation: See "Off screen capture" on page 100
ACQuire:ROLLmode:POINts?
Returns the record length of the roll mode. In roll mode, the complete record is always
captured, independently of the displayed waveform. To process and analyze the com-
plete roll waveform, enable ACQuire:ROLLmode:OSCapture.
Return values:
<RecordLength> Range: 0 to 18446744073709551615
Increment: 1
*RST: 0
Default unit: pts
Usage: Query only
Asynchronous command
Manual operation: See "Off screen capture" on page 100
TIMebase:ROLL:STATe?
Returns the status of the roll mode.
Return values:
<State> ON | OFF
*RST: OFF
Usage: Query only
Asynchronous command
Manual operation: See "Mode" on page 100
17.8.4 Vertical setup
The channel suffix <ch> selects the input channel that is affected by the command.
CHANnel<ch>:STATe..................................................................................................... 430
CHANnel<ch>:SCALe.................................................................................................... 430
CHANnel<ch>:RANGe................................................................................................... 430
CHANnel<ch>:OFFSet................................................................................................... 431
Acquisition and setup
Loading ...
Loading ...
Loading ...