Remote control commands
R&S
®
MXO 4 Series
432User Manual 1335.5337.02 ─ 05
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Value> DC | DCLimit | AC
DC
Connection with 50 Ω termination, passes both DC and AC com-
ponents of the signal.
DCLimit
Connection with 1 MΩ termination, passes both DC and AC
components of the signal.
AC
Connection with 1 MΩ termination through DC capacitor,
removes DC and very low-frequency components. The wave-
form is centered on zero volts.
Usage:
Asynchronous command
Manual operation: See "Coupling" on page 113
CHANnel<ch>:INVert <InvertChannel>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the volt-
age values of all signal components against the ground level.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<InvertChannel> ON | OFF
ON: inverted waveform
OFF: normal waveform
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Invert channel" on page 114
CHANnel<ch>:SKEW:TIME <Offset>
Sets a skew value to compensate for the delay of the measurement setup or from the
circuit specifics that the instrument cannot compensate automatically. It affects only the
selected input channel.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Offset> Range: -1E-07 to 1E-07
Increment: 1E-12
*RST: 0
Default unit: s