Remote control commands
R&S
®
MXO 4 Series
471User Manual 1335.5337.02 ─ 05
SHORter
Triggers if the signal crosses vertical limits before the specified
width time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
width time is reached.
*RST: WITHin
Usage: Asynchronous command
Manual operation: See "Time condition" on page 148
TRIGger:EVENt<ev>:WINDow:WIDTh
<Width>
Depending on the values of TRIGger:EVENt<ev>:WINDow:RANGe the width sets:
For the ranges Within and Outside, the width defines the center of a time range.
The range is defined by the limits "±Delta", see TRIGger:EVENt<ev>:WINDow:
DELTa.
For the ranges Shorter and Longer, it defines the maximum and minimum time
lapse, respectively.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Width> Range: 1E-10 to 10000
Increment: 1E-07
*RST: 5E-09
Default unit: s
Usage: Asynchronous command
Manual operation: See "Width" on page 148
17.9.8 Timeout trigger
TRIGger:EVENt<ev>:TIMeout:RANGe............................................................................. 471
TRIGger:EVENt<ev>:TIMeout:TIME.................................................................................472
TRIGger:EVENt<ev>:TIMeout:RANGe <TimeoutMode>
Sets the relation of the signal level to the trigger level for the timeout trigger.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Trigger