Rohde & Schwarz MXO44-2410 Oscilloscope

User Manual - Page 510

For MXO44-2410.

PDF File Manual, 759 pages, Read Online | Download pdf file

MXO44-2410 photo
Loading ...
Loading ...
Loading ...
Remote control commands
R&S
®
MXO 4 Series
510User Manual 1335.5337.02 ─ 05
COUPled
The current horizontal settings of the diagram are used.
*RST: ORIGinal
Usage: Asynchronous command
Manual operation: See "Mode" on page 191
REFCurve<rc>:RESCale:HORizontal:FACTor <ScaleFactor>
Sets the horizontal scale factor. A factor greater than 1 stretches the waveform hori-
zontally, a factor lower than 1 compresses the curve.
Suffix:
<rc>
.
1...4, index of the reference waveform
Parameters:
<ScaleFactor> Range: 1E-06 to 1000000
Increment: 0.1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Scale factor" on page 191
REFCurve<rc>:RESCale:HORizontal:OFFSet <Offset>
Moves the waveform horizontally. Enter a value with a time unit suitable for the time
scale of the diagram.
Positive values shift the waveform to the right, negative values shift it to the left.
Suffix:
<rc>
.
1...4, index of the reference waveform
Parameters:
<Offset> Range: -1E+26 to 1E+26
Increment: 0.01
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Horizontal offset" on page 191
REFCurve<rc>:RESCale:HORizontal:STATe <State>
If enabled, the horizontal offset and factor are applied to the reference waveform.
Stretching and offset change the display of the waveform independent of the horizontal
settings of the source waveform and of the horizontal diagram settings.
Suffix:
<rc>
.
1...4, index of the reference waveform
Waveform analysis
Loading ...
Loading ...