Remote control commands
R&S
®
MXO 4 Series
438User Manual 1335.5337.02 ─ 05
Usage: Asynchronous command
Manual operation: See "Interpolation" on page 106
ACQuire:POINts[:VALue] <RecordLength>
Sets the record length in manual record length mode. In automatic record length mode,
it sets the maximum value of the record length for automatic calculation.
The record length is the number of waveform samples that are stored in one waveform
record after processing, including interpolation.
Parameters:
<RecordLength> Range: 1000 to Device specific
Increment: 2
*RST: 1000
Default unit: pts
Usage: Asynchronous command
Manual operation: See "Record length, Record length limit" on page 105
ACQuire:POINts:ARATe?
Returns the sample rate of the ADC, before waveform processing. The result is the
interleaved sample rate or the non-interleaved one, depending on the channel usage.
Return values:
<ADCSampleRate> Range: 5 Gsample/s (interleaved), 2,5 Gsample/s (non-
interleaved)
Default unit: Hz
Usage: Query only
Asynchronous command
ACQuire:POINts:MAXimum <RecLengthLimit>
Sets the maximum record length, if ACQuire:POINts:MODE is se to AUTO.
Parameters:
<RecLengthLimit> Range: 1000 to 800E+6
Increment: 2
*RST: 10E+6
Default unit: pts
Usage: Asynchronous command
ACQuire:POINts:MODE <RecLengthMode>
Selects the mode of the waveform record length adjustment.
Parameters:
<RecLengthMode> AUTO | MANual