5.12 Channel Delay
When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring relatively greater errors (zero offset). This series oscilloscope
allows you to set a delay time for calibrating the zero offset of the corresponding
channel. Zero offset is defined as the offset of the crossing point of the waveform and
trigger level line relative to the trigger position, as shown in
Figure 5.2
.
Figure 5.2 Zero Offset
Press > More > CH-CH Skew, rotate the multifunction knob or use the
numeric keypad to set the desired delay calibration time. The available range of the
delay calibration time is from -100 ns to 100 ns.
NOTE
This parameter is related to the instrument model and the current horizontal time base setting.
The larger the horizontal time base is, the larger the setting step will be.
5.13 To Set the Bias
When you use an oscilloscope to make actual measurements, a small offset that arises
from the temperature drift of the component or external environment disturbance
may occur on the zero-cross voltage of the channel, which will affect the
measurement results of the vertical parameters. This series oscilloscope allows you to
set an offset calibration voltage for calibrating the zero point of the corresponding
channel, so as to improve the accuracy of the measurement results.