1. Delay A →B : indicates the time difference between the threshold middle values
on the rising edge of Source A and that of Source B. Negative delay indicates that
the rising edge of Source A occurred after that of Source B.
2. Delay A →B : indicates the time difference between the threshold middle values
on the falling edge of Source A and that of Source B. Negative delay indicates that
the falling edge of Source A occurred after that of Source B.
3. Delay A →B : indicates the time difference between the threshold middle values
on the rising edge of Source A and the falling edge of Source B. Negative delay
indicates that the rising edge of Source A occurred after the falling edge of Source
B.
4. Delay A →B : indicates the time difference between the threshold middle values
on the falling edge of Source A and the rising edge of Source B. Negative delay
indicates that the falling edge of Source A occurred after the rising edge of Source
B.
5. Phase A →B : indicates the phase deviation between the threshold middle
values on the rising edge of Source A and that of Source B. The phase formula is as
follows:
°360×=
sourceA
RR
RR
Period
BDelayA
BPhaseA
Wherein,
RR
BPhaseA
represents Phase A →B ,
RR
BDelayA
represents Delay A
→B , and
sourceA
Period
represents period of Source A.
6. Phase A →B : indicates the phase deviation between the threshold middle
values on the falling edge of Source A and that of Source B. The phase formula is
as follows:
°360×=
sourceA
FF
FF
Period
BDelayA
BPhaseA
Wherein,
FF
BPhaseA
represents Phase A →B ,
FF
BDelayA
represents Delay A
→B , and
sourceA
Period
represents period of Source A.