keys are illuminated for a few seconds. The oscilloscope starts to perform a series of
self-tests. After the self-test, the splash screen is displayed.
NOTE
Restart and Shutdown:
1. Enable the touch screen and then tap the function navigation icon at the lower-left
corner of the touch screen to enable the function navigation.
2. Tap the "Restart" icon, and then you can select "Restart" or "Shutdown", as shown in the
figure below.
3. If you tap "Restart" , the oscilloscope powers off and then automatically restarts again.
If you tap "Shutdown"
, the oscilloscope powers off (you can also press the Power key
at the lower-left corner of the front panel to turn off the oscilloscope).
4.3.4 Fuse Replacement
If you need to replace the fuse, please use the specified fuse (AC 250 V, T3.15 A; 5.2
mm × 20 mm) and perform the following operations (as shown in
Figure 4.5
):
1. Turn off the instrument, cut off the power, and remove the power cord.
2. Use a small straight slotted screwdriver to pry out the fuse holder.
3. Take out the fuse holder.
4. Replace the old fuse with a specified fuse.
5. Install the fuse holder.