Press More > Sample Position, then rotate the multifunction knob or use the
pop-up numeric keypad to set the sample position. The settable range is from 10% to
90%.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press to quickly switch
the current trigger mode. For details, refer to descriptions in "
Trigger Mode
".
Trigger Parameter Setting:
Sets the trigger parameter (noise rejection) under this trigger type. For details, refer
to descriptions in "
Noise Rejection
".
Trigger LEVEL/Threshold Level:
Rotate the Trigger LEVEL knob to adjust the trigger level/ threshold level. Please
refer to descriptions in "
Trigger LEVEL/Threshold Level
". The current trigger level/
threshold level is displayed at the upper-right corner of the screen.
8.7.17 FlexRay Trigger (Option)
This series oscilloscope can trigger on the specified frame, symbol, error, or position
of the FlexRay bus.
FlexRay is a type of differential serial bus configured with three consecutive segments
(i.g. packet header, payload, and packet trailer). Its data transmission rate is up to 10
Mb/s. Each frame contains a static segment and a dynamic segment, and ends with
the bus idle time.
Its format is as shown in the figure below.
Figure 8.20 Frame Format of FlexRay Bus
Trigger Type:
Press Type, and then rotate the multifunction knob to select "FlexRay". Press down
the knob to select the trigger type. Then, the current trigger setting information is
To Trigger the Oscilloscope
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
121