7. EXT TRIG
Inputs the external signal via the external input terminal.
8. Trigger Out and Pass/Fail
- TRIG OUT
The oscilloscope can output a signal that can reflect the current capture rate of
the oscilloscope at each trigger via this interface. Connect the signal to a
waveform display device and measure the frequency of the signal. The
measurement result is the same as the current capture rate.
- Pass/Fail
The instrument can output a pulse from the [TRIG OUT] connector when a
pass/failed event is detected during the pass/fail test.
9. Lock Hole
You can lock the instrument to a fixed location by using the security lock (please
purchase it by yourself) via the lock hole.
10. Fuse
If you need to replace the fuse, use only the specified fuse. For details, refer to
descriptions in "
Fuse Replacement
".
11. AC Power Cord Connector
Indicates the input terminal where AC power source is connected. The power
supply requirements of the instrument are: 100 V~240 V; 45 Hz~ 440 Hz. Please
use the power cord provided in the accessories to connect the oscilloscope to the
AC power source.
4.6 Front Panel Function Overview
4.6.1 Vertical
• , , , : indicates the analog channel switch key. The four channels are
marked by different colors which are also used to mark both the corresponding
Quick Start
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
25