Measurements
R&S
®
RTC1000
86User Manual 1335.7352.02 ─ 04
If deactivated, the cursors stay in position on the screen if the waveform is scaled, and
the measured value changes.
7.2.2 Measurement Types
The following cursor measurement types are available:
Voltage
"Voltage" provides two cursors and measures three voltages. The values V1 and V2
correspond to the voltage between the zero base line of the selected signal and the
current position of the first or second cursor. The value ΔV is voltage difference
between the two cursors.
Time
"Time" provides two cursors and measures three times and an equivalent frequency.
The values t1 and t2 correspond to the time between the trigger and the current posi-
tion of the first or second cursor. The value Δt corresponds to the time between the two
cursors.
Ratio X
"Ratio X" provides three cursors and measures a ratio in X direction (e.g. a duty ratio)
between the first two cursors and the first and the third cursor. The measurement val-
ues are displayed in four formats (floating point, percent, degrees and radians).
Ratio Y
"Ratio Y" provides three cursors and measures a ratio in Y direction (e.g. an over-
shoot) between the first two cursors and the first and the third cursor. The measure-
ment values are displayed in two formats (floating point and percent).
Count
"Count" provides three cursors to count signal changes that exceed the threshold
within a specific interval. The interval is set by the first two cursors and the threshold is
the third cursor. The measurement values are displayed in four formats (number of ris-
ing and falling slopes and number of positive and negative pulses).
Peak Values
"Peak Values" provides two cursors and measures the minimum and the maximum
voltage of a signal within the interval that is set by both cursors. The values Vp- and Vp
+ correspond to the minimum and the maximum voltage. The peak value (Vpp) corre-
sponds to the voltage difference between the minimum and maximum value.