Rohde & Schwarz RTC1002EDU oscilloscope

Rohde & Schwarz RTC1000 manual - Page 70

For RTC1002EDU.

PDF File Manual, 371 pages, Read Online | Download pdf file

RTC1002EDU photo
Loading ...
Loading ...
Loading ...
Analysis
R&S
®
RTC1000
70User Manual 1335.7352.02 ─ 04
3. Select "EDIT" = "Display".
4. Assign the unit "A" (Ampere) to the equation by pressing "UNIT".
5. Press "LABEL" and enter CURRENT.
6. Select "EDIT" = "Parameter".
7. Select "CH1" as "OPERAND 1" and "Const." as "OPERAND 2".
8. Press "EDIT CONSTANT" to define the constant: "NUMERIC INPUT" = 0.1.
Figure 6-10: Definition of the current equation
The correct scaling and the units for cursors and automatic measurements are
defined.
To define the POWER equation
1. Add a new equation "MA2" to formulary set 1.
2. Select "EDIT" = "Parameter".
3. Select "Multiplication" as "OPERATOR", "CURRENT" as "OPERAND 1" and "CH2"
as "OPERAND 2".
4. Select "EDIT" = "Display".
5. Assign the unit "W" (Watt) to the equation.
6. Press t"LABEL" and enter POWER.
Mathematics
Loading ...
Loading ...
Loading ...