Waveform Setup
R&S
®
RTC1000
38User Manual 1335.7352.02 ─ 04
Make sure to align passive probes before their first use, after an extended measure-
ment break, or when switching instruments or channels.
Two connector pins for compensation are located at the front panel. The left pin is on
ground level. The next pin supplies a square wave signal for the adjustment.
To perform probe compensation using the wizard
1. Press the [SETUP] key in the General section.
2. Press the "PROBE ADJUST" softkey.
3. Follow the instructions of the wizard. It guides you through the compensation proc-
ess.
4. Use the compensation trimmer of the probe to get optimum square wave response.
For details, refer to the documentation of your probe.
The compensation is performed at 1 kHz (LF) and 1 MHz (HF).